The GlideScope® Video Laryngoscope: randomized clinical trial in 200 patients
The GlideScope® Video Laryngoscope is a new intubating device. It was designed to provide a view of the glottis without alignment of the oral, pharyngeal and tracheal axes. The aim of the study was to describe the use of the GlideScope® in comparison with direct laryngoscopy for elective surgical pa...
Gespeichert in:
Veröffentlicht in: | British journal of anaesthesia : BJA 2005-03, Vol.94 (3), p.381-384 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The GlideScope® Video Laryngoscope is a new intubating device. It was designed to provide a view of the glottis without alignment of the oral, pharyngeal and tracheal axes. The aim of the study was to describe the use of the GlideScope® in comparison with direct laryngoscopy for elective surgical patients requiring tracheal intubation.
Two hundred patients were randomly assigned to intubation by direct laryngoscopy using a Macintosh size 3 blade (DL, n=100) or intubation using the GlideScope® (GS, n=100). Prior to intubation all patients were given a Cormack and Lehane (C&L) grade by a separate anaesthetist using a Macintosh size 3 blade. The patient was then intubated, using direct laryngoscopy or the GlideScope®, by a different anaesthetist during which the larynx was inspected and given a laryngoscopy score. Time to intubate was measured.
In the GS group, laryngoscopy grade was improved in the majority (28/41) of patients with C&L grade >1 and in all but one of patients who were grade 3 laryngoscopy (P |
---|---|
ISSN: | 0007-0912 1471-6771 |