Parameters involved in the conversion of real-time PCR assays from the ABI prism 7700 to the Cepheid SmartCycler® II

We examined several critical parameters that must be optimized when converting between the ABI Prism 7700 real-time PCR platform and the Cepheid SmartCycler® II while using the same primer and probe sequences. A lyophilized master mix, MgCl 2 concentration, PCR cycling conditions, and ramp times wer...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical biochemistry 2005-02, Vol.38 (2), p.183-186
Hauptverfasser: Bentley, Heather A., Belloni, Dorothy R., Tsongalis, Gregory J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We examined several critical parameters that must be optimized when converting between the ABI Prism 7700 real-time PCR platform and the Cepheid SmartCycler® II while using the same primer and probe sequences. A lyophilized master mix, MgCl 2 concentration, PCR cycling conditions, and ramp times were evaluated. Optimization of each parameter, including use of the OmniMix™ HS-lyophilized beads, 6 mM MgCl 2 concentration, changes in PCR cycling parameters, and increased ramp time were necessary to convert this real time PCR assay to a new platform. We conclude that careful consideration of several analytical parameters can result in a smooth transition of assays between real time PCR platforms.
ISSN:0009-9120
1873-2933
DOI:10.1016/j.clinbiochem.2004.10.009