Parameters involved in the conversion of real-time PCR assays from the ABI prism 7700 to the Cepheid SmartCycler® II
We examined several critical parameters that must be optimized when converting between the ABI Prism 7700 real-time PCR platform and the Cepheid SmartCycler® II while using the same primer and probe sequences. A lyophilized master mix, MgCl 2 concentration, PCR cycling conditions, and ramp times wer...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2005-02, Vol.38 (2), p.183-186 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We examined several critical parameters that must be optimized when converting between the ABI Prism 7700 real-time PCR platform and the Cepheid SmartCycler® II while using the same primer and probe sequences.
A lyophilized master mix, MgCl
2 concentration, PCR cycling conditions, and ramp times were evaluated.
Optimization of each parameter, including use of the OmniMix™ HS-lyophilized beads, 6 mM MgCl
2 concentration, changes in PCR cycling parameters, and increased ramp time were necessary to convert this real time PCR assay to a new platform.
We conclude that careful consideration of several analytical parameters can result in a smooth transition of assays between real time PCR platforms. |
---|---|