Cementless Porous-Coated Sockets Without Holes Implanted With Pure Press-Fit Technique
We evaluated the use of a pure press-fit technique in 220 porous-coated acetabular components without screw holes at mean follow-up of 6.1 years (4-7.5 years). The socket had a 14° enhanced gradient at the periphery and was underreamed by 1 to 2 mm depending on bone quality. The Harris Hip average s...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2005, Vol.20 (1), p.4-10 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated the use of a pure press-fit technique in 220 porous-coated acetabular components without screw holes at mean follow-up of 6.1 years (4-7.5 years). The socket had a 14° enhanced gradient at the periphery and was underreamed by 1 to 2 mm depending on bone quality. The Harris Hip average score was 44.3 (26-70) preoperatively and 93.1 (66-100) at final follow-up. There were no postoperative gaps in any zone. Nonprogressive radiolucent lines were detected in 4 hips. There was no movement or migration of the acetabular component on the follow-up x-rays. Pelvic osteolysis was not observed. Whereas 5 cups (2.3%) had revisions, none was revised for aseptic loosening. The press-fit technique provides stability in appropriate cases avoiding the theoretical complications related to screws or screw holes. |
---|---|