Compression forces generated by Mini bone screws—a comparative study done on bone model
The compressive forces generated by the AO/ASIF 3.0 mm cannulated cancellous and 2.0 mm cortical screws, Mini-Acutrak and Herbert/Whipple small bone cannulated screws were measured in the laboratory with the use of simulated cancellous bone and a load cell washer as a means of quantifying their fixa...
Gespeichert in:
Veröffentlicht in: | Injury 2005, Vol.36 (1), p.65-70 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The compressive forces generated by the AO/ASIF 3.0
mm cannulated cancellous and 2.0
mm cortical screws, Mini-Acutrak and Herbert/Whipple small bone cannulated screws were measured in the laboratory with the use of simulated cancellous bone and a load cell washer as a means of quantifying their fixation capabilities. The Herbert/Whipple screw and the Mini-Acutrak screw were found to have nearly identical compression capabilities and provided more compression than the cortical screw. The AO/ASIF cannulated screw when used with a support screw demonstrated a compressive capacity twice that of the 2.0
mm cortical screw and higher than the headless Mini-Acutrak and Herbert/Whipple screws. The Mini-Acutrak screw produced about 70% of compression of the cancellous screw in spite of having a diameter almost half that of the cancellous screw. The Herbert/Whipple screw in spite of its larger size compared to the Mini-Acutrak produced almost the same amount of compression. |
---|---|
ISSN: | 0020-1383 1879-0267 |
DOI: | 10.1016/j.injury.2004.03.026 |