Glucose Sensor Based on Glucose Oxidase Immobilized by Zirconium Phosphate
Amperometric glucose sensors were fabricated using glucose oxidase (GOx) entrapped in zirconium hydrogenphosphate (ZrP), and their performance was evaluated. Reportedly, α-ZrP is one of the candidates that are expected to improve the stability of enzymes immobilized on solid surfaces.1,2 We intercal...
Gespeichert in:
Veröffentlicht in: | Analytical Sciences 2004, Vol.20(12), pp.1635-1638 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Amperometric glucose sensors were fabricated using glucose oxidase (GOx) entrapped in zirconium hydrogenphosphate (ZrP), and their performance was evaluated. Reportedly, α-ZrP is one of the candidates that are expected to improve the stability of enzymes immobilized on solid surfaces.1,2 We intercalated GOxs into ZrP (GOx/ZrP), cast the GOx/ZrP suspension in polyvinylalcohol on a platinum electrode, and dried it in a vacuum oven. The morphological layered structure was investigated by scanning electron microscopy. The enzymatic activities, which were determined by open-circuit potentiometric technique, reached the highest when GOxs were immobilized in ZrP at ca. pH 5. In vitro tests showed good linear responses in the 0 - 25 mM range and the sensitivity of 0.14 nA mM-1 at 0.4 V vs. Ag/AgCl. The sensors, as made, were stable for more than 3 days within a limited deterioration. |
---|---|
ISSN: | 0910-6340 1348-2246 |
DOI: | 10.2116/analsci.20.1635 |