Potential Primary Measurement Tool for the Quantification of DNA

An automated sample introduction system, utilizing a demountable direct injection high-efficiency nebulizer (d-DIHEN), is successfully incorporated for the first time with an inductively coupled plasma optical emission spectrometer (ICP-OES) for the measurement of the phosphorus content in acid-dige...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 2009-05, Vol.81 (9), p.3414-3420
Hauptverfasser: Brennan, Ryan G, Rabb, Savelas A, Holden, Marcia J, Winchester, Michael R, Turk, Gregory C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An automated sample introduction system, utilizing a demountable direct injection high-efficiency nebulizer (d-DIHEN), is successfully incorporated for the first time with an inductively coupled plasma optical emission spectrometer (ICP-OES) for the measurement of the phosphorus content in acid-digested nucleotides and deoxyribonucleic acid (DNA). With this experimental setup, the solution uptake rate and volume are reduced from 170 to 30 μL min−1 and from 10 to 2.4 mL, respectively, thereby reducing the required DNA sample mass for solutions containing 3 μg g−1 P from 300 to 72 μg of DNA, in comparison to previous analyses in our laboratory using a glass concentric nebulizer with cyclonic spray chamber arrangement. The use of direct injection also improves P (I) 213.617 nm sensitivity by a factor of 4 on average. A high-performance (HP) methodology in combination with the previous sample introduction system and ICP-OES provides simultaneous, time-correlated internal standardization and drift correction resulting in relative expanded uncertainties (% U) for the P mass fractions in the range of 0.1−0.4 (95% confidence level) for most of the thymidine 5′-monophosphate (TMP), calf thymus DNA (CTDNA), and plasmid DNA (PLDNA) analyses. The d-DIHEN with HP-ICP-OES methodology allows for the quantification of DNA mass at P mass fractions as low as 0.5 μg g−1, further reducing the required DNA mass to 12 μg, with small uncertainty (≤0.4%). This successful approach will aid in the development and certification of nucleic acid certified reference materials (CRMs), particularly for these samples that are typically limited in volume.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac802688x