Evaluation of nano-liquid chromatography-tandem mass spectrometry in a column switching setup for the absolute quantification of peptides in the picomolar range

A standard nanospray-liquid chromatography-tandem mass spectrometry system in a column switching setup for the absolute quantification of leucine-enkephalin was evaluated. Analytes were loaded on a C18 trapping column and back-flushed in the 75 microm analytical column. Quantification was performed...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Chromatography A 2004-11, Vol.1058 (1-2), p.113-119
Hauptverfasser: Sinnaeve, Bart A, Van Bocxlaer, Jan F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A standard nanospray-liquid chromatography-tandem mass spectrometry system in a column switching setup for the absolute quantification of leucine-enkephalin was evaluated. Analytes were loaded on a C18 trapping column and back-flushed in the 75 microm analytical column. Quantification was performed with a triple quadrupole instrument. Validation results show that it is feasible, with a conventional nano-LC system in the column switching setup, to quantify peptides as low as 500 amol on column (50 pmol/L). Weighted linear regression analysis proves a good linearity in a dynamic range of almost three orders of magnitude. Nevertheless, robustness remains a key issue in nano-LC-MS/MS.
ISSN:0021-9673
DOI:10.1016/S0021-9673(04)01304-4