Evaluation of nano-liquid chromatography-tandem mass spectrometry in a column switching setup for the absolute quantification of peptides in the picomolar range
A standard nanospray-liquid chromatography-tandem mass spectrometry system in a column switching setup for the absolute quantification of leucine-enkephalin was evaluated. Analytes were loaded on a C18 trapping column and back-flushed in the 75 microm analytical column. Quantification was performed...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 2004-11, Vol.1058 (1-2), p.113-119 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A standard nanospray-liquid chromatography-tandem mass spectrometry system in a column switching setup for the absolute quantification of leucine-enkephalin was evaluated. Analytes were loaded on a C18 trapping column and back-flushed in the 75 microm analytical column. Quantification was performed with a triple quadrupole instrument. Validation results show that it is feasible, with a conventional nano-LC system in the column switching setup, to quantify peptides as low as 500 amol on column (50 pmol/L). Weighted linear regression analysis proves a good linearity in a dynamic range of almost three orders of magnitude. Nevertheless, robustness remains a key issue in nano-LC-MS/MS. |
---|---|