Low background shielding of HPGe detector
National Radiation Protection Institute in Prague is equipped with 14 HPGe detectors with relative efficiency up to 150%. Steel shielding with one of these detectors (relative efficiency 100%) was chosen to be rebuilt to decrease minimum detectable activity (MDA). Additional lead and copper shieldin...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2009-05, Vol.67 (5), p.723-725 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | National Radiation Protection Institute in Prague is equipped with 14 HPGe detectors with relative efficiency up to 150%. Steel shielding with one of these detectors (relative efficiency 100%) was chosen to be rebuilt to decrease minimum detectable activity (MDA). Additional lead and copper shielding was built up inside the original steel shielding to reduce the volume of the inner space and filled with nitrogen by means of evaporating liquid nitrogen. MDA values decreased for Compton background up to 0.67 of original value. |
---|---|