Volume estimation of small phantoms and rat kidneys using three-dimensional ultrasonography and a position sensor
To evaluate the accuracy of small volume estimation, both in vivo and in vitro, measurements with a three-dimensional (3D) ultrasound (US) system were carried out. A position sensor was used and the transmitting frequency was 10 MHz. Balloons with known volumes were scanned while rat kidneys were sc...
Gespeichert in:
Veröffentlicht in: | Ultrasound in medicine & biology 2004-09, Vol.30 (9), p.1109-1117 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the accuracy of small volume estimation, both
in vivo and
in vitro, measurements with a three-dimensional (3D) ultrasound (US) system were carried out. A position sensor was used and the transmitting frequency was 10 MHz. Balloons with known volumes were scanned while rat kidneys were scanned
in vivo and
in vitro. The Archimedes’ principle was used to estimate the true volume. For balloons, the 3D US system gave very good agreement with true volumes in the volume range 0.1 to 10.0 mL (r = 0.999,
n = 45, mean difference ± 2SD = 0.245 ± 0.370 mL). For rat kidneys
in vivo (volume range 0.6 to 2.7 mL) the method was less accurate (r = 0.800,
n = 10, mean difference ± 2SD = −0.288 ± 0.676 mL). For rat kidneys
in vitro (volume range 0.3 to 2.7 mL) the results showed good agreement (r = 0.981,
n = 23, mean difference ± 2SD = 0.039 ± 0.254 mL). For balloons, kidneys
in vivo and
in vitro, the mean percentage error was 9.3 ± 4.8%, −17.1 ± 17.4%, and 4.6 ± 11.5%, respectively. This method can estimate the volume of small phantoms and rat kidneys and opens new possibilities for volume measurements of small objects and the study of organ function in small animals. (E-mail