Effectiveness and safety of the Proxis system in demonstrating retrograde coronary blood flow during proximal occlusion and in capturing embolic material
The Feasibility And Safety Trial for its embolic protection device during transluminal intervention in coronary vessels: a European Registry (FASTER) was designed to demonstrate that (1) the Proxis embolic protection system can control anterograde flow and reverse blood flow in native coronary arter...
Gespeichert in:
Veröffentlicht in: | The American journal of cardiology 2004-11, Vol.94 (9), p.1134-1139 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Feasibility And Safety Trial for its embolic protection device during transluminal intervention in coronary vessels: a European Registry (FASTER) was designed to demonstrate that (1) the Proxis embolic protection system can control anterograde flow and reverse blood flow in native coronary arteries and saphenous vein grafts; and (2) this system can capture embolic debris. Percutaneous coronary intervention on stenotic coronary lesions revolutionized treatment of coronary disease, but is associated with the risk of major adverse cardiac events. This prospective, nonrandomized, multicenter clinical feasibility and safety study enrolled 40 patients with 51 lesions at 4 centers who underwent treatment of stenotic lesions with proximal emboli protection (Proxis system). Proxis was successfully used 95% of the time, and embolic debris was qualitatively identified in all cases. Major adverse cardiac events occurred in 2 patients (5.0%): 1 late in-stent thrombosis resulting in death and 1 non–Q-wave infarction when a lesion was crossed before deployment of the Proxis system. Mean vessel occlusion time was 4.3 ± 2.4 minutes. Native flow reversal was sufficient in 31 patients, with a mean aspirate volume of 11.8 ± 6.5 ml. When the infusion catheter was used to augment reflow, mean occlusion time was 4.6 minutes. In conclusion, this trial is the first to demonstrate that retrograde blood flow can be achieved during proximal occlusion and that the Proxis system can be used safely during intervention of saphenous vein grafts and native coronary arterial lesions to capture embolic material. |
---|---|
ISSN: | 0002-9149 1879-1913 |
DOI: | 10.1016/j.amjcard.2004.07.079 |