Highly selective preconcentration of Cu(II) from seawater and water samples using amidoamidoxime silica
Chemically modified silica containing amidoamidoxime group was studied as a sorbent for solid-phase extraction (SPE) and preconcentration of Cu(II) prior to determination by flame atomic absorption spectrometry (FAAS). The sorbent showed an extremely high selectivity towards Cu(II) in the pH range o...
Gespeichert in:
Veröffentlicht in: | Talanta (Oxford) 2009-05, Vol.78 (3), p.1004-1010 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Chemically modified silica containing amidoamidoxime group was studied as a sorbent for solid-phase extraction (SPE) and preconcentration of Cu(II) prior to determination by flame atomic absorption spectrometry (FAAS). The sorbent showed an extremely high selectivity towards Cu(II) in the pH range of 4–6, while the extraction of Pb(II), Cd(II), Ni(II) and Co(II) was low. The adsorption isotherm followed the Langmuir model and the maximum sorption capacity of 0.0163
mmol
Cu(II)
g
−1 was achieved. In the flow system, Cu(II) was completely retained on a column containing 40
mg of the modified silica at the flow rate of 4.0
mL
min
−1 and quantitatively eluted by 5
mL of 1% (v/v) HNO
3. No interference from Na
+, K
+, Mg
2+, Ca
2+, Cl
− and SO
4
2− at 10, 100 and 1000
mg
L
−1 was observed. When applied for preconcentration and determination of Cu(II) in tap water, pond water, and seawater, the recoveries were 96, 101, and 95%, respectively, with high precision (% relative standard deviation (R.S.D.)
<
4) and low method detection limit (9
μg
L
−1). |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/j.talanta.2009.01.017 |