Iterative reverse optimization procedure for calibration of aspheric wave-front measurements on a nonnull interferometer

The accuracy of interferometric measurements made with a nonnull configuration is degraded by test-dependent aberrations. Calibration of the data can be done with reverse optimization methods. An iterative reverse optimization process that improves efficiency and eliminates sensitive merit function...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 2004-09, Vol.43 (27), p.5152-5161
Hauptverfasser: Gappinger, Robert O, Greivenkamp, John E
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The accuracy of interferometric measurements made with a nonnull configuration is degraded by test-dependent aberrations. Calibration of the data can be done with reverse optimization methods. An iterative reverse optimization process that improves efficiency and eliminates sensitive merit function weighting issues is described. The process is shown to calibrate a nonnull interferometric measurement of a wave front with more than 200 waves of departure to an accuracy of 0.16 waves peak to valley and 0.02 waves rms.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/AO.43.005152