Management of gamma nail breakage with bipolar hemi-arthroplasty

Gamma nail breakage is an uncommon occurrence that often arises from fatigue failure of the implant, with a reported incidence ranging from 0.2 to 5.7 percent. We report a 73-year-old woman with a three-part intertrochanteric fracture and who presented two years postoperatively with gamma nail failu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Singapore medical journal 2009-01, Vol.50 (1), p.e44-e47
Hauptverfasser: Wee, J L H, Sathappan, S S, Yeo, M S W, Low, Y P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Gamma nail breakage is an uncommon occurrence that often arises from fatigue failure of the implant, with a reported incidence ranging from 0.2 to 5.7 percent. We report a 73-year-old woman with a three-part intertrochanteric fracture and who presented two years postoperatively with gamma nail failure secondary to fracture non-union. This patient underwent a revision long-stem bipolar hemi-arthroplasty and has been followed-up for 24 months, with good functional and radiological results.
ISSN:0037-5675