Computer-assisted surgery for dynamic hip screw, using Surgix©, a novel intraoperative guiding system

Background We present Surgix©, a new computer‐assisted surgery system (CASS). This system uses image analysis technology in order to measure three‐dimensional (3D) distances, visualize implant templates and view the guided trajectory on standard fluoroscopy. Our purpose was to compare surgery result...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The international journal of medical robotics + computer assisted surgery 2009-03, Vol.5 (1), p.45-50
Hauptverfasser: Herman, Amir, Dekel, Assaf, Botser, Itamar B., Steinberg, Ely L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background We present Surgix©, a new computer‐assisted surgery system (CASS). This system uses image analysis technology in order to measure three‐dimensional (3D) distances, visualize implant templates and view the guided trajectory on standard fluoroscopy. Our purpose was to compare surgery results and technique with and without the Surgix CASS. Methods The study included 61 dynamic hip screw (DHS) procedures. The Surgix system was used in 41 procedures. We compared the number of guide wire insertion trials and the time needed for each trial, the number of X‐ray pulses, tip–apex index, nine‐quadrant position and shaft–neck angle. Results The procedures were carried out by experienced users (≥ five operations, using the system) and had a first‐trial guide wire insertion success rate of 77.8%, compared to a rate of 10% for the control group (p = 0.001) and fewer insertion trials (1.33 vs. 3.05, respectively; p = 0.001). The mean number of fluoroscopy pulses was 41.5% lower for the experienced group than for the control (17.6 vs. 30.1; p = 0.009). There were no significant differences in tip–apex distance, favourable quadrant screw placement or neck–shaft angle. Conclusion The results of this study demonstrate that the Surgix CASS significantly improves the accuracy of hardware positioning and reduces radiation exposure time, thus enhancing patient outcome. Copyright © 2008 John Wiley & Sons, Ltd.
ISSN:1478-5951
1478-596X
DOI:10.1002/rcs.231