Severe Valvular Regurgitation and Late Prosthesis Embolization After Percutaneous Aortic Valve Implantation
We report the case of a 79-year-old man with low-flow, low-gradient aortic stenosis who underwent transapical aortic valve implantation. The deployment of a 26-mm Edwards SAPIEN valve (Edwards Lifesciences Inc, Irvine, CA) was complicated with the occurrence of severe central aortic regurgitation th...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 2009-02, Vol.87 (2), p.618-621 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report the case of a 79-year-old man with low-flow, low-gradient aortic stenosis who underwent transapical aortic valve implantation. The deployment of a 26-mm Edwards SAPIEN valve (Edwards Lifesciences Inc, Irvine, CA) was complicated with the occurrence of severe central aortic regurgitation that was successfully treated with the implantation of a second valve (“valve-in-valve”). Postoperative progress was satisfactory but 2 days after the procedure he suddenly developed cardiogenic shock secondary to the migration of the aortic bioprothesis into the left ventricular outflow tract. Open-heart surgery was undertaken to remove the valves and perform standard aortic valve replacement, but the patient died perioperatively. |
---|---|
ISSN: | 0003-4975 1552-6259 |