Laserinterferometric assessment of pilocarpine-induced movement of an accommodating intraocular lens: A randomized trial
To measure the axial movement of an accommodating intraocular lens (IOL) induced by ciliary muscle contraction after application of pilocarpine. Randomized, controlled, patient- and examiner-masked trial with intrapatient comparison. One hundred ten eyes of 55 patients with age-related bilateral cat...
Gespeichert in:
Veröffentlicht in: | Ophthalmology (Rochester, Minn.) Minn.), 2004-08, Vol.111 (8), p.1515-1521 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To measure the axial movement of an accommodating intraocular lens (IOL) induced by ciliary muscle contraction after application of pilocarpine.
Randomized, controlled, patient- and examiner-masked trial with intrapatient comparison.
One hundred ten eyes of 55 patients with age-related bilateral cataract.
This study was divided into 3 parts. In the first, the accommodating IOL (1CU) was compared with a 3-piece open-loop acrylic IOL that served as the control. In the second, to assess the effect of capsule fibrosis on the potential accommodating performance of the accommodating IOL, extensive polishing of the anterior capsule with a slit cannula was compared with standard surgery. In the third, the effect of a posterior capsulorhexis was compared with that of standard surgery. Anterior chamber depth (ACD) was assessed with partial coherence interferometry, measured before and after topical application of pilocarpine 2%, and near visual acuity (VA) was evaluated 3 months after surgery.
Pilocarpine-induced change in ACD.
The accommodating IOL showed a forward movement under pilocarpine with a median amplitude of movement of −314 μm (95% confidence interval [CI]: −148 to −592), compared with the backward movement of 63 μm (95% CI: 161 to −41) for the open-loop control IOL (
P = 0.001). Capsule polishing and a posterior capsulorhexis had no effect on IOL movement with the accommodating IOL. The median near VA with distance correction was 20/60.
Pilocarpine induced a small but significant forward movement of the accommodating IOL. However, the amount of movement was calculated to result in a refractive change of |
---|---|