Evaluation of an Adaptive Directional System in a DSP Hearing Aid

The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of audiology 2004-06, Vol.13 (1), p.73-79
Hauptverfasser: Bentler, Ruth A, Tubbs, Jill L, Egge, Jessica L. M, Flamme, Gregory A, Dittberner, Andrew B
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (cardioid) setting and the adaptive setting. Five speakers were placed between 110 degrees and 250 degrees azimuth around the listener. Speech-weighted noise was presented from those speakers at an overall level (OAL) of 65 dB (A). Noise was increased by 8 dB from 1 speaker at a time, using 2-s modulation and random assignment, while the output from the other speakers was reduced to maintain the constant OAL. Results of 2 speech perception tasks used as outcome measures indicated that the adaptive system was not able to follow the dominant noise source in the presence of lower level noise sources. Self-report measures obtained after blinded home trials were consistent with laboratory findings that the participants did not perceive this adaptive microphone design to be more effective than the default fixed-microphone option.
ISSN:1059-0889
1558-9137