Evaluation of an Adaptive Directional System in a DSP Hearing Aid
The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (...
Gespeichert in:
Veröffentlicht in: | American journal of audiology 2004-06, Vol.13 (1), p.73-79 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (cardioid) setting and the adaptive setting. Five speakers were placed between 110 degrees and 250 degrees azimuth around the listener. Speech-weighted noise was presented from those speakers at an overall level (OAL) of 65 dB (A). Noise was increased by 8 dB from 1 speaker at a time, using 2-s modulation and random assignment, while the output from the other speakers was reduced to maintain the constant OAL. Results of 2 speech perception tasks used as outcome measures indicated that the adaptive system was not able to follow the dominant noise source in the presence of lower level noise sources. Self-report measures obtained after blinded home trials were consistent with laboratory findings that the participants did not perceive this adaptive microphone design to be more effective than the default fixed-microphone option. |
---|---|
ISSN: | 1059-0889 1558-9137 |
DOI: | 10.1044/1059-0889(2004/010) |