Comparison of percutaneous closure of large patent ductus arteriosus by multiple coils versus the Amplatzer duct occluder device

This study compared the efficacy and costs of the most used approaches for percutaneous closure of large patent ductus arteriosus, that is, multiple coils and the Amplatzer duct occluder (ADO) device. From April 2000 to September 2003, 47 patients underwent closure of large, symptomatic patent ductu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The American journal of cardiology 2004-07, Vol.94 (2), p.252-255
Hauptverfasser: Santoro, Giuseppe, Bigazzi, Maurizio Cappelli, Palladino, Maria Teresa, Russo, Maria Giovanna, Carrozza, Marianna, Calabrò, Raffaele
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study compared the efficacy and costs of the most used approaches for percutaneous closure of large patent ductus arteriosus, that is, multiple coils and the Amplatzer duct occluder (ADO) device. From April 2000 to September 2003, 47 patients underwent closure of large, symptomatic patent ductus arteriosus (diameter 4.6 ± 3.0 mm/m 2; QP/QS 2.1 ± 1.9) with multiple Cook detachable coils (n = 19) or the ADO device (n = 28). The multiple coil approach was significantly cheaper (1,389 ± 168 vs 3,811 ± 38 €, p
ISSN:0002-9149
1879-1913