Multiple-wavelength reference based on interleaved, sampled fiber Bragg gratings and molecular absorption

We present a wavelength calibration reference based on interleaved, sampled fiber Bragg gratings stabilized to a molecular absorption line. Such a hybrid reference can provide multiple stable calibration peaks over a wide range of wavelengths. We demonstrate a wavelength reference that has at least...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 2004-06, Vol.43 (17), p.3530-3534
Hauptverfasser: Rowe, Mary A, Swann, William C, Gilbert, Sarah L
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present a wavelength calibration reference based on interleaved, sampled fiber Bragg gratings stabilized to a molecular absorption line. Such a hybrid reference can provide multiple stable calibration peaks over a wide range of wavelengths. We demonstrate a wavelength reference that has at least 20 peaks suitable for use as calibration references in each of three wavelength regions: 850, 1300, and 1550 nm. We monitored the stability of a 1300-nm reflection peak and found that the standard deviation of the peak wavelength was 0.7 pm over a 70-day period.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/AO.43.003530