Assessment of Stenosis in Vascular Access Grafts
: The major complication that occurs with grafts used as vascular access for hemodialysis, is stenosis at the venous anastomosis or in the draining vein. 75% area stenosis is considered significant as thrombotic occlusion may occur. The aim of this experimental study was to evaluate invasive and no...
Gespeichert in:
Veröffentlicht in: | Artificial organs 2004-07, Vol.28 (7), p.617-622 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | : The major complication that occurs with grafts used as vascular access for hemodialysis, is stenosis at the venous anastomosis or in the draining vein. 75% area stenosis is considered significant as thrombotic occlusion may occur. The aim of this experimental study was to evaluate invasive and noninvasive indices to detect significant stenoses in a vascular access graft. A compliant underarm loop graft in vitro model was built and studied with 50, 65, 80, and 90% stenosis at flow rates of 500, 1000, and 1500 mL/min. Flow in the system was pulsatile. Velocity was measured with ultrasound Doppler and the pressure was measured invasively. The resistance index (RI), pvenous line/MAP, and the newly introduced pressure ratio (PR) were calculated and compared. A stenosis can be suspected when a high frequency ultrasound velocity signal develops at the venous anastomosis. RI > 1 confirms a very severe stenosis (90%). The parameter PR |
---|---|
ISSN: | 0160-564X 1525-1594 |
DOI: | 10.1111/j.1525-1594.2004.07368.x |