Development of indirect-competitive quartz crystal microbalance immunosensor for C-reactive protein
A highly sensitive indirect-competitive quartz crystal microbalance immunosensor for a cardiovascular biomarker, C-reactive protein (CRP), was developed. The sensor system was constructed with a buffer reservoir, a micro-dispensing pump, an injector, a flow-through cell with a sensor chip, an oscill...
Gespeichert in:
Veröffentlicht in: | Sensors and actuators. B, Chemical Chemical, 2009-12, Vol.143 (1), p.444-448 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A highly sensitive indirect-competitive quartz crystal microbalance immunosensor for a cardiovascular biomarker, C-reactive protein (CRP), was developed. The sensor system was constructed with a buffer reservoir, a micro-dispensing pump, an injector, a flow-through cell with a sensor chip, an oscillator module, a frequency response analyzer and a PC. The reaction buffer for the system was 0.1
M sodium phosphate buffer (pH 7.0). The optimum concentrations of CRP for immobilization and an anti-CRP antibody for competition reaction were 2 and 0.250
mg/mL, respectively. When plotted in double-logarithmic scale, a linear relationship between CRP concentration and sensor response was obtained in a wide concentration range of 0.130–25,016
ng/mL of CRP, together with the limit of detection of 0.130
ng/mL. |
---|---|
ISSN: | 0925-4005 1873-3077 |
DOI: | 10.1016/j.snb.2009.09.046 |