Development of the Sealed-Type Triple-Point-of-Argon Cell for Long-Stem SPRT Calibration at KRISS
A system was fabricated to realize the triple point of argon for the calibration of long-stem standard platinum resistance thermometers. A cryostat was constructed so that the temperature could be controlled quasi-adiabatically, and the melting was realized using the continuous-heating method. The c...
Gespeichert in:
Veröffentlicht in: | International journal of thermophysics 2008-10, Vol.29 (5), p.1740-1748 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A system was fabricated to realize the triple point of argon for the calibration of long-stem standard platinum resistance thermometers. A cryostat was constructed so that the temperature could be controlled quasi-adiabatically, and the melting was realized using the continuous-heating method. The combined uncertainty of the realization of the triple point of argon for a confidence level of 95% was 0.6 mK. |
---|---|