Development of the Sealed-Type Triple-Point-of-Argon Cell for Long-Stem SPRT Calibration at KRISS

A system was fabricated to realize the triple point of argon for the calibration of long-stem standard platinum resistance thermometers. A cryostat was constructed so that the temperature could be controlled quasi-adiabatically, and the melting was realized using the continuous-heating method. The c...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of thermophysics 2008-10, Vol.29 (5), p.1740-1748
Hauptverfasser: Yang, I., Song, C. H., Kang, K. H., Kim, Y. -G., Gam, K. S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A system was fabricated to realize the triple point of argon for the calibration of long-stem standard platinum resistance thermometers. A cryostat was constructed so that the temperature could be controlled quasi-adiabatically, and the melting was realized using the continuous-heating method. The combined uncertainty of the realization of the triple point of argon for a confidence level of 95% was 0.6 mK.
ISSN:0195-928X
1572-9567