Effect of Beam Size on Stone Comminution in Shock Wave Lithotripsy
The effect of beam size on stone comminution in the Dornier HM-3 lithotripter were investigated using a modified reflector (MR), which was developed to produce an acoustic field with higher peak pressure and smaller beam size compared to the original reflector (OR) of the HM-3. The acoustic fields p...
Gespeichert in:
Hauptverfasser: | , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The effect of beam size on stone comminution in the Dornier HM-3 lithotripter were investigated using a modified reflector (MR), which was developed to produce an acoustic field with higher peak pressure and smaller beam size compared to the original reflector (OR) of the HM-3. The acoustic fields produced at 20 kV by the MR in the focal plane, and by the OR in the focal and pre-focal plane (z = -15 mm) were characterized using a light spot hydrophone. The efficiencies of stone comminution in a mesh holder were similar, but the corresponding values produced by using the MR in a membrane holder were significantly lower than those produced by the OR in the focal and pre-focal planes. These results suggest that a broad beam size could increase stone comminution efficiency when fragments are spreading out or moving due to respiratory motion in a large area during SWL. In contrast, when stone fragments are confined and well aligned to lithotripter focus the beam size may not influence significantly the treatment outcome. |
---|---|
ISSN: | 0094-243X |
DOI: | 10.1063/1.3131443 |