Performance of a helium circulation system for a MEG

We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cryogenics (Guildford) 2009-03, Vol.49 (3), p.144-150
Hauptverfasser: Takeda, Tsunehiro, Okamoto, Masayoshi, Atsuda, Kazuhiro, Katagiri, Keishi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without any noise problems. The noise level is below 10 fT/Hz 1/2 for 2–40 Hz, below 30 fT/Hz 1/2 at 1 Hz, and 200 fT/Hz 1/2 at 50 Hz, which is the power supply frequency. As the amount of liquid helium used decreases less than one percent, the maintenance cost of the MEG becomes less than one-tenth of the previous cost.
ISSN:0011-2275
1879-2235
DOI:10.1016/j.cryogenics.2008.12.001