Performance of a helium circulation system for a MEG
We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without a...
Gespeichert in:
Veröffentlicht in: | Cryogenics (Guildford) 2009-03, Vol.49 (3), p.144-150 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5
W GM cryocoolers operating at 4.2
K. The MEG with the HCS was used to measure human brain responses for over one and a half years without any noise problems. The noise level is below 10 fT/Hz
1/2 for 2–40
Hz, below 30 fT/Hz
1/2 at 1
Hz, and 200 fT/Hz
1/2 at 50
Hz, which is the power supply frequency. As the amount of liquid helium used decreases less than one percent, the maintenance cost of the MEG becomes less than one-tenth of the previous cost. |
---|---|
ISSN: | 0011-2275 1879-2235 |
DOI: | 10.1016/j.cryogenics.2008.12.001 |