Realization of the Triple Point of Water in Metallic Sealed Cells at the LNE-INM/CNAM: A Progress Report
A miniature metallic cell for the water triple point (TPW, temperature 273.16 K) was developed for capsule-type thermometer calibrations for realizations with adiabatic calorimetry techniques. The LNE-INM/Cnam previously developed a copper cell for the water triple point and the techniques for clean...
Gespeichert in:
Veröffentlicht in: | International journal of thermophysics 2008-06, Vol.29 (3), p.825-835 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A miniature metallic cell for the water triple point (TPW, temperature 273.16 K) was developed for capsule-type thermometer calibrations for realizations with adiabatic calorimetry techniques. The LNE-INM/Cnam previously developed a copper cell for the water triple point and the techniques for cleaning, filling, and sealing. On the basis of previous work, a new copper cell prototype for the TPW was developed and filled at the LNE-INM/Cnam. Measurements were performed using an appropriate calorimeter and a comparison block containing several thermometers. Preliminary results show a scatter of the temperatures measured at the phase transition of the order of 0.2 mK when measurements are repeated over a short-term period (1 month). A positive drift in the phase transition temperature of about 30μK·month
−1
was observed over several months. Studies are in progress to improve the cell, to reduce the reproducibility uncertainty to less than 0.1 mK and to have a phase transition with better temporal stability. |
---|---|
ISSN: | 0195-928X 1572-9567 |