Intercomparison of the Realizations of the ITS-90 from 83.8058 K to 692.677 K among European NMIs
The EUROMET.T-K3 comparison is the regional extension of CCT-K3. The comparison involved the six European national metrology institutes (NMIs) previously involved in CCT-K3 (LNE-INM/CNAM, SMU, INRiM, NMi-VSL, NPL, PTB) and 18 additional European national laboratories. The comparison was divided into...
Gespeichert in:
Veröffentlicht in: | International journal of thermophysics 2008-06, Vol.29 (3), p.991-1000 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The EUROMET.T-K3 comparison is the regional extension of CCT-K3. The comparison involved the six European national metrology institutes (NMIs) previously involved in CCT-K3 (LNE-INM/CNAM, SMU, INRiM, NMi-VSL, NPL, PTB) and 18 additional European national laboratories. The comparison was divided into five different loops, each coordinated by a co-pilot chosen from the laboratories having participated in the CCT-K3 comparison. LNE-INM/CNAM played the role of pilot in linking the five loops. In each loop, an artifact in the form of a standard platinum resistance thermometer (SPRT, 25 Ω) was circulated among the participating laboratories. To have sufficient information about the possible drift of the SPRTs, the co-pilots performed a calibration over the full temperature range at the beginning and at the end of the loop. A EUROMET reference value (ERV), taking into account the whole comparison, was defined, and the differences (
T
Lab
−
T
ERV
) were calculated with the associated uncertainties. The method for establishing the link between the participants in CCT-K3 and in EUROMET.T-K3 is described. |
---|---|
ISSN: | 0195-928X 1572-9567 |
DOI: | 10.1007/s10765-008-0397-x |