Experimental fluid mechanics of pulsatile artificial blood pumps

The fluid mechanics of artificial blood pumps has been studied since the early 1970s in an attempt to understand and mitigate hemolysis and thrombus formation by the device. Pulsatile pumps are characterized by inlet jets that set up a rotational “washing” pattern during filling. Strong regurgitant...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annual review of fluid mechanics 2006-01, Vol.38 (1), p.65-86
Hauptverfasser: DEUTSCH, Steven, TARBELL, John M, MANNING, Keefe B, ROSENBERG, Gerson, FONTAINE, Arnold A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The fluid mechanics of artificial blood pumps has been studied since the early 1970s in an attempt to understand and mitigate hemolysis and thrombus formation by the device. Pulsatile pumps are characterized by inlet jets that set up a rotational “washing” pattern during filling. Strong regurgitant jets through the closed artificial heart valves have Reynolds stresses on the order of 10,000 dynes/cm 2 and are the most likely cause of red blood cell damage and platelet activation. Although the flow in the pump chamber appears benign, low wall shear stresses throughout the pump cycle can lead to thrombus formation at the wall of the smaller pumps (10–50 cc). The local fluid mechanics is critical. There is a need to rapidly measure or calculate the wall shear stress throughout the device so that the results may be easily incorporated into the design process.
ISSN:0066-4189
1545-4479
DOI:10.1146/annurev.fluid.38.050304.092022