Evaluation of the applicability of urine lateral flow immunochromatography tests for the detection of cocaine in plasma samples

•Benzoylecgonine was verified as the best target analyte, with a cutoff of 40 ng/mL.•Benzoylecgonine at 40 ng/mL presented 96.3 % sensitivity and 97.9 % specificity.•Comparison of immunoassay results with cocaine levels was poor.•The use of plasma in the urine cocaine test proved to be adequate and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical biochemistry 2025-01, Vol.135, p.110854, Article 110854
Hauptverfasser: de Souza, Isabella Almeida Millan, dos Santos, Bruno Pereira, do Nascimento, Sabrina Nunes, Birk, Letícia, Sebben, Viviane Cristina, Eller, Sarah, de Oliveira, Tiago Franco
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•Benzoylecgonine was verified as the best target analyte, with a cutoff of 40 ng/mL.•Benzoylecgonine at 40 ng/mL presented 96.3 % sensitivity and 97.9 % specificity.•Comparison of immunoassay results with cocaine levels was poor.•The use of plasma in the urine cocaine test proved to be adequate and reliable. Qualitative analysis of cocaine in urine is a common practice in emergency settings. However, positive results from urine screening tests do not necessarily indicate recent exposure. In this context, plasma is considered a more appropriate option due to its shorter detection window and better correlation with symptomatology. Therefore, the availability of rapid tests for this biological matrix is extremely relevant in the clinical and emergency context. A lateral flow immunochromatography test designed for analyzing cocaine in urine was evaluated for use with plasma samples. A total of 412 samples from suspected cases of intoxication were processed and tested. Concurrently, the samples were analyzed for cocaine and its metabolites, benzoylecgonine and ecgonine methyl ester, using liquid chromatography-tandem mass spectrometry (LC-MS/MS). Reliability parameters, including sensitivity, specificity, positive predictive value, negative predictive value, and efficiency, were calculated considering different cutoff values. Approximately 8.3 % of the samples tested positive in the immunoassay, while 10.2 % had a concentration greater than 5 ng/mL for at least one analyte in the LC-MS/MS analysis. Using benzoylecgonine as the target analyte with a cutoff of 40 ng/mL yielded the best reliability results, with 96.3 % sensitivity and 97.9 % specificity. Cocaine did not show satisfactory results, whereas ecgonine methyl ester, despite having 92.9 % sensitivity and 94.7 % specificity at its best cutoff (20 ng/mL), had a positive predictive value of only 38.2 %. The study evaluated the suitability of using rapid urine tests for cocaine detection in plasma, offering a simple and quick drug screening method that is particularly useful in toxicological emergencies.
ISSN:0009-9120
1873-2933
1873-2933
DOI:10.1016/j.clinbiochem.2024.110854