Evaluation of a novel robotic testing method for stability and kinematics of total knee arthroplasty
This work developed a novel preclinical test of total knee replacements (TKRs) in order to explain TKR instability linked to patient dissatisfaction. It was hypothesized that stability tests on the isolated moving prostheses would provide novel comparative data on the stability and kinematics among...
Gespeichert in:
Veröffentlicht in: | Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA sports traumatology, arthroscopy : official journal of the ESSKA, 2024-10 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This work developed a novel preclinical test of total knee replacements (TKRs) in order to explain TKR instability linked to patient dissatisfaction. It was hypothesized that stability tests on the isolated moving prostheses would provide novel comparative data on the stability and kinematics among TKR designs.
Three TKR designs, DePuy Synthes Attune MS, Stryker Triathlon and Zimmer Biomet Persona MC, were assessed using a robotic arm while flexing-extending 0-140°. Tests imposed 710 N body weight combined with three tibial loads: no anterior-posterior (AP) force, 90 N anterior or 90 N posterior force. Other load effects were minimized and the kinematics was recorded. Each implant was tested six times to investigate the repeatability of the method. Data were analysed using statistical parametric mapping with one-way analysis of variance (ANOVA). If significance was found (p |
---|---|