Endovascular Conversion of a Failed Nellix AAA-Repair by a Custom-Made Branched Device
This technical note presents a case of a patient with a failed Nellix device (Endologix, Irvine, Calif) who was not deemed fit for open conversion. Our planned approach for repair involved an endovascular procedure utilizing a custom-made branched device. An endovascular repair was performed via a c...
Gespeichert in:
Veröffentlicht in: | Journal of endovascular therapy 2024-08, p.15266028241276961 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This technical note presents a case of a patient with a failed Nellix device (Endologix, Irvine, Calif) who was not deemed fit for open conversion. Our planned approach for repair involved an endovascular procedure utilizing a custom-made branched device.
An endovascular repair was performed via a custom-made four outer branched device in conjunction with a custom-made bifurcated graft featuring inverted limbs (Cook Inc., Bloomington, Ind). All branches were connected to the target vessel with Gore Viabahn VBX balloon-expandable covered stents (Gore & Associates Inc.).
Endovascular conversion with branched endovascular repair in a patient not deemed fit for open surgery was successfully performed, thereby reducing the risk of a high morbid and mortal procedure.
Novel useful treatment solution of failed EVAR with Nellix device. |
---|---|
ISSN: | 1526-6028 1545-1550 1545-1550 |