The Impact of Dose Calibration Point Differences Between the Treatment Plan and Patient-specific Quality Assurance in Passive Scattering Proton Beam Therapy
Passive scattering proton beam therapy (PSPT) is performed by taking actual measurements of all pre-designated fields in a treatment plan followed by appropriate adjustments to the prescribed dose. For this reason, it is necessary to ensure precision management of the measurements (patient-specific...
Gespeichert in:
Veröffentlicht in: | In vivo (Athens) 2024-09, Vol.38 (5), p.2478-2483 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Passive scattering proton beam therapy (PSPT) is performed by taking actual measurements of all pre-designated fields in a treatment plan followed by appropriate adjustments to the prescribed dose. For this reason, it is necessary to ensure precision management of the measurements (patient-specific calibration) in the administration of a planned dose. Therefore, this study investigated the impact on dose distribution in treatment planning when the patient calibration point differs from the normalized point in a treatment plan.
A total of 16 cases were selected, where the patient calibration point and normalized point did not match, and the normalized point used in the treatment plan was changed to the patient calibration point using a treatment planning system (VQA ver. 2.01, HITACHI). At this point, the displacement of the relative dose at the isocenter was estimated as an error owing to the difference compared to the patient calibration point.
Overall, the error was within the range of ±1.5%, with the exception of orbit cases. Calibrated points also tended to be lower than the normalized points in the treatment plan. In terms of treatment sites, a greater deviation was observed for head cases. Cases with a large deviation in sites other than the head were attributed to poor flatness within the radiation field owing to a narrower opening of the patient collimator.
Dose measurement errors in PSPT due to differing calibration points were generally within ±1.5%, with higher deviations observed in head treatments because of complex structures and narrow collimator openings. A γ analysis for significant deviations showed a 98.7% passing rate, suggesting limited overall impact. It is important to select stable calibration points in dosimetry to ensure high precision in dose administration, particularly in complex treatment areas. |
---|---|
ISSN: | 0258-851X 1791-7549 1791-7549 |
DOI: | 10.21873/invivo.13718 |