Ischemic core detection threshold of computed tomography perfusion (CTP) in acute stroke
Purpose This study aimed to determine the accuracy of detecting ischemic core volume using computed tomography perfusion (CTP) in patients with suspected acute ischemic stroke compared to diffusion-weighted magnetic resonance imaging (DW-MRI) as the reference standard. Methods This retrospective mon...
Gespeichert in:
Veröffentlicht in: | Radiologia medica 2024-10, Vol.129 (10), p.1522-1529 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
This study aimed to determine the accuracy of detecting ischemic core volume using computed tomography perfusion (CTP) in patients with suspected acute ischemic stroke compared to diffusion-weighted magnetic resonance imaging (DW-MRI) as the reference standard.
Methods
This retrospective monocentric study included patients who underwent CTP and DW-MRI for suspected acute ischemic stroke. The ischemic core size was measured at DW-MRI. The detectability threshold volume was defined as the lowest volume detected by each method. Clinical data on revascularization therapy, along with the clinical decision that influenced the choice, were collected. Volumes of the ischemic cores were compared using the Mann–Whitney
U
test.
Results
Of 83 patients who underwent CTP, 52 patients (median age 73 years, IQR 63–80, 36 men) also had DW-MRI and were included, with a total of 70 ischemic cores. Regarding ischemic cores, only 18/70 (26%) were detected by both CTP and DW-MRI, while 52/70 (74%) were detected only by DW-MRI. The median volume of the 52 ischemic cores undetected on CTP (0.6 mL, IQR 0.2–1.3 mL) was significantly lower (