The temperature profile in elbow arthroscopy using radiofrequency ablation

Introduction Over the last decades, arthroscopic surgery has become increasingly relevant as its minimally invasive approach offers many benefits. To investigate the risks of orthoscopic surgery at the elbow, this study aimed to investigate the development of temperatures in elbow joints while perfo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Archives of orthopaedic and trauma surgery 2024-08, Vol.144 (8), p.3517-3522
Hauptverfasser: Krane, Felix, Wegmann, Kilian, Kahmann, Stephanie, Harbrecht, Andreas, Peter, Lars, Leschinger, Tim
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction Over the last decades, arthroscopic surgery has become increasingly relevant as its minimally invasive approach offers many benefits. To investigate the risks of orthoscopic surgery at the elbow, this study aimed to investigate the development of temperatures in elbow joints while performing radiofrequency ablation in arthroscopic surgery. Materials and methods We performed standard arthroscopic surgeries with posterolateral, transtricipital and anterolateral approaches on seven cadaveric elbows and performed ablation on predefined locations with or without irrigation. Two temperature probes were positioned into the olecranon fossa and between the ulnar nerve and the medial joint capsule. The temperature data were recorded using a real-time data logger. A bipolar radiofrequency ablation (bRFA) device was used at the medial and lateral recess and in the fossa olecrani over a defined period. Data was then analyzed using Matlab. Results Using bRFA without irrigation results in rapidly increasing temperature within the joint. A significant temperature increase was found within only 5 s without irrigation ( p  = 0.0052) in the fossa olecrani. We did not observe critical temperatures above 41 °C close to the ulnar nerve within 30 s under constant irrigation ( p  = 0.0747). Conclusions Radiofrequency ablation (RFA) can be safely used in elbow arthroscopy with irrigation. The continuous use without irrigation should be limited to 3 s. Despite the anatomical proximity of the ulnar nerve and capsule, we were able to show that a possible rise in temperature most likely does not affect the ulnar nerve.
ISSN:1434-3916
0936-8051
1434-3916
DOI:10.1007/s00402-024-05472-6