Ultra‐low‐field magnetization transfer imaging at 0.055T with low specific absorption rate
Purpose To demonstrate magnetization transfer (MT) effects with low specific absorption rate (SAR) on ultra‐low‐field (ULF) MRI. Methods MT imaging was implemented by using sinc‐modulated RF pulse train (SPT) modules to provide bilateral off‐resonance irradiation. They were incorporated into 3D grad...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2024-12, Vol.92 (6), p.2420-2432 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To demonstrate magnetization transfer (MT) effects with low specific absorption rate (SAR) on ultra‐low‐field (ULF) MRI.
Methods
MT imaging was implemented by using sinc‐modulated RF pulse train (SPT) modules to provide bilateral off‐resonance irradiation. They were incorporated into 3D gradient echo (GRE) and fast spin echo (FSE) protocols on a shielding‐free 0.055T head scanner. MT effects were first verified using phantoms. Brain MT imaging was conducted in both healthy subjects and patients.
Results
MT effects were clearly observed in phantoms using six SPT modules with total flip angle 3600° at central primary saturation bands of approximate offset ±786 Hz, even in the presence of large relative B0 inhomogeneity. For brain, strong MT effects were observed in gray matter, white matter, and muscle in 3D GRE and FSE imaging using six and sixteen SPT modules with total flip angle 3600° and 9600°, respectively. Fat, cerebrospinal fluid, and blood exhibited relatively weak MT effects. MT preparation enhanced tissue contrasts in T2‐weighted and FLAIR‐like images, and improved brain lesion delineation. The estimated MT SAR was 0.0024 and 0.0008 W/kg for two protocols, respectively, which is far below the US Food and Drug Administration (FDA) limit of 3.0 W/kg.
Conclusion
Robust MT effects can be readily obtained at ULF with extremely low SAR, despite poor relative B0 homogeneity in ppm. This unique advantage enables flexible MT pulse design and implementation on low‐cost ULF MRI platforms to achieve strong MT effects in brain and beyond, potentially augmenting their clinical utility in the future. |
---|---|