Accuracy of intraocular lens power formulas in patients with average keratometry greater than 46 diopters
To compare the accuracy of Kane, Barrett Universal II, Haigis, and SRK-T formulas in eyes with average keratometry greater than 46 diopters (D). A retrospective analysis was conducted on 101 eyes of 101 patients with average keratometry greater than 46 D. The absolute prediction error (EA) was obtai...
Gespeichert in:
Veröffentlicht in: | Archivos de la Sociedad Española de Oftalmología (English ed.) 2024-10, Vol.99 (10), p.413-419 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To compare the accuracy of Kane, Barrett Universal II, Haigis, and SRK-T formulas in eyes with average keratometry greater than 46 diopters (D).
A retrospective analysis was conducted on 101 eyes of 101 patients with average keratometry greater than 46 D. The absolute prediction error (EA) was obtained for each patient one month after surgery. The mean absolute prediction error (MEA), median absolute prediction error (MedEA) and the percentage of patients with absolute refractive error less than 0.25 D, 0.50 D, and 1.00 D were calculated for each formula analyzed.
The Kane formula achieved the lowest MEA (0.53 ± 0.43) and the lowest MedEA (0.41), followed by Barrett Universal II (MEA: 0.56 ± 0.42, MedEA: 0.49), SRK-T (MEA: 0.59 ± 0.44, MedEA: 0.54), and Haigis (MEA: 0.77 ± 0.47, MedEA: 0.69), showing a significant difference in the results. It was also observed that the Kane formula was the most accurate, with the highest percentage of patients, with EA less than 0.25 D, 0.50 D, and 1.00 D (30.7%, 54.4%, and 86.1%, respectively), while the Haigis formula was the least accurate (12.9%, 33.7%, and 69.3%, respectively).
In eyes with corneas having average keratometry greater than 46 D, the Kane formula proves to be a useful tool in intraocular lens (IOL) power calculation and demonstrates higher precision compared to the Barrett Universal II, SRK-T, and Haigis formulas.
Comparar la precisión de las fórmulas Kane, Barrett Universal II, Haigis y SRK-T en ojos con queratometrías promedio mayores a 46 dioptrías (D).
Se analizaron de forma retrospectiva 101 ojos de 101 pacientes con queratometrías promedio mayores a 46 D, se obtuvo el error de predicción absoluto (EA) de cada paciente al mes de su cirugía y se calcularon la media del error de predicción absoluto (MEA), la mediana del error de predicción absoluto (MedEA) y el porcentaje de pacientes con EA menor a 0.25 D, 0.50 D y 1.00 D para cada una de las fórmulas a analizar.
La fórmula de Kane obtuvo la menor MEA (0.53 ± 0.43) y la menor MedEA (0.41), seguida de Barrett Universal II (MEA: 0.56 ± 0.42, MedAE: 0.49), SRK-T (MEA: 0.59 ± 0.44, MedEA: 0.54) y Haigis (MEA: 0.77 ± 0.47, MedEA: 0.69), mostrando una diferencia significativa en el resultado. Se observó también que la fórmula Kane fue la más precisa, obteniendo el mayor porcentaje de pacientes con EA menor a 0.25 D, 0.50 D y 1.00 D (30.7%, 54.4% y 86.1% respectivamente) y la fórmula Haigis la menos precisa (12.9%, 33.7% y 69.3% respectivamente).
En ojos c |
---|---|