Development, validation and long-term evaluation of a liquid chromatography-tandem mass spectrometry method for simultaneous quantification of amiodarone, desethylamiodarone and mexiletine in human plasma and serum

•Quantification of amiodarone, desethylamiodarone and mexiletine in one single run.•Efficient and robust LC-MS/MS method for guiding anti-arrhythmic therapy in ventricular arrhythmia.•The validated method was applied and evaluated in clinical practice for therapeutic drug monitoring. Amiodarone and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2024-08, Vol.1243, p.124233, Article 124233
Hauptverfasser: Braakhuis, Martinus W.A., Pistorius, Marcel C.M., Postema, Pieter G., Hollak, Carolina E.M., Swart, Eleonora L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•Quantification of amiodarone, desethylamiodarone and mexiletine in one single run.•Efficient and robust LC-MS/MS method for guiding anti-arrhythmic therapy in ventricular arrhythmia.•The validated method was applied and evaluated in clinical practice for therapeutic drug monitoring. Amiodarone and mexiletine are used for ventricular arrhythmias, for which a combination therapy of both anti-arrhythmic drugs (AADs) is not uncommon. Therapeutic drug monitoring (TDM) can be beneficial for clinical guidance of therapy, especially to correctly identify adverse events. Desethylamiodarone, the active metabolite of amiodarone, accumulates over time and is associated with serious adverse events. Therefore, simultaneous TDM for amiodarone, desethylamiodarone and mexiletine is advantageous in clinical practice. The presented LC-MS/MS method was validated for selectivity, matrix effect, linearity, accuracy, precision, carry-over and stability. The method was continuously evaluated during eight months of clinical use. The method was shown to be linear within the measured range of 0.1 to 10 mg/L for each component. The matrix effect was considered negligible. No interfering responses were found for amiodarone, desethylamiodarone and the isotopic-labeled internal standards. A constant and reproducible within-run contribution of 45.3 %, originating from the system, was identified for mexiletine. The systemic contribution to the peak area of the lowest quantifiable concentration of mexiletine affected the selectivity and carry-over effect measurements. Multiple measurements showed that regression adjusted concentrations were accurate and reproducible, indicating calibration correction was applicable. Sample stability was found to be within limits for all storage conditions and freeze–thaw cycles. Furthermore, long-term method evaluation with external controls resulted in stable measurements with a percentage coefficient of variance between 1.3 % and 6.3 %. The presented practical and reliable method is applicable for clinical TDM and will allow clinical practitioners to guide drug therapy of amiodarone and mexiletine.
ISSN:1570-0232
1873-376X
1873-376X
DOI:10.1016/j.jchromb.2024.124233