APTT critical value establishment in four different reagent/instrument systems based on single factor deficiencies
Introduction There are significant differences in the activated partial thromboplastin time (APTT) critical values reported in different studies, most of which does not make recommendations for any specific clear detection systems. The International Council for Standardization in Hematology (ICSH) r...
Gespeichert in:
Veröffentlicht in: | Annals of hematology 2024-08, Vol.103 (8), p.3219-3227 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
There are significant differences in the activated partial thromboplastin time (APTT) critical values reported in different studies, most of which does not make recommendations for any specific clear detection systems. The International Council for Standardization in Hematology (ICSH) recommends that APTT critical values be established based on the reagent type, coagulation factor sensitivity and heparin response. The objective of this study was to establish APTT critical values by using different reagents and based on single coagulation factor deficiencies.
Methods
The APTT values were determined in commercial endogenous coagulation factor-deficient plasma at concentrations of 1 IU/dL, 2 IU/dL, 5 IU/dL, 10 IU/dL, 20 IU/dL, and 30 IU/dL by using four assay systems. The retrospective collection of data from patients who lacked factor VIII (FVIII), FIX, or FXI alone was performed. Receiver operating characteristic (ROC) curves were constructed to assess the diagnostic accuracy of APTT for identifying patients with an endogenous coagulation factor activity |
---|---|
ISSN: | 0939-5555 1432-0584 1432-0584 |
DOI: | 10.1007/s00277-024-05718-8 |