Point-of-Care Bedside Brain Magnetic Resonance Imaging Is Safe in Extracorporeal Membrane Oxygenation Patients With Swan Ganz Catheters: A Phantom Experiment and Single Center Experience

More than 1.2 million pulmonary artery catheters (PACs) are used in cardiac patients per annum within the United States. However, it is contraindicated in traditional 1.5 and 3T magnetic resonance imaging (MRI) scans. We aimed to test preclinical and clinical safety of using this imaging modality gi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of surgical research 2024-07, Vol.299, p.290-297
Hauptverfasser: Chinedozi, Ifeanyi D., Boskamp, Eddy, Darby, Zachary, Kang, Jin Kook, Rando, Hannah, Sair, Haris, Pitt, John, Wilcox, Christopher, Kim, Bo Soo, Khanduja, Shivalika, Whitman, Glenn, Cho, Sung-Min
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:More than 1.2 million pulmonary artery catheters (PACs) are used in cardiac patients per annum within the United States. However, it is contraindicated in traditional 1.5 and 3T magnetic resonance imaging (MRI) scans. We aimed to test preclinical and clinical safety of using this imaging modality given the potential utility of needing it in the clinical setting. We conducted two phantom experiments to ensure that the electromagnetic field power deposition associated with bare and jacketed PACs was safe and within the acceptable limit established by the Food and Drug Administration. The primary end points were the safety and feasibility of performing Point-of-Care (POC) MRI without imaging-related adverse events. We performed a preclinical computational electromagnetic simulation and evaluated these findings in nine patients with PACs on veno-arterial extracorporeal membrane oxygenation. The phantom experiments showed that the baseline point specific absorption rate through the head averaged 0.4 W/kg. In both the bare and jacketed catheters, the highest net specific absorption rates were at the neck entry point and tip but were negligible and unlikely to cause any heat-related tissue or catheter damage. In nine patients (median age 66, interquartile range 42-72 y) with veno-arterial extracorporeal membrane oxygenation due to cardiogenic shock and PACs placed for close hemodynamic monitoring, POC MRI was safe and feasible with good diagnostic imaging quality. Adult ECMO patients with PACs can safely undergo point-of-care low-field (64 mT) brain MRI within a reasonable timeframe in an intensive care unit setting to assess for acute brain injury that might otherwise be missed with conventional head computed tomography.
ISSN:0022-4804
1095-8673
1095-8673
DOI:10.1016/j.jss.2024.04.045