Accurate detection of lead malfunction from ECG-derived bipolar pacing stimulus amplitude

One common mode of lead failure is insulation breach, which may result in myopotential noise and device malfunction. “Pseudo-unipolarization” of bipolar pacing stimuli, as observed from a routine 12-lead electrocardiogram (ECG) due to stimulus current leak, has been observed with insulation breaches...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Heart rhythm 2024-11, Vol.21 (11), p.2226-2231
Hauptverfasser: Lloyd, Michael S., Pelling, Mary, Ibrahim, Rand, El-Chami, Mikhael F., Iravanian, Shahriar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:One common mode of lead failure is insulation breach, which may result in myopotential noise and device malfunction. “Pseudo-unipolarization” of bipolar pacing stimuli, as observed from a routine 12-lead electrocardiogram (ECG) due to stimulus current leak, has been observed with insulation breaches. We sought to characterize this electrocardiographic finding to detect this type of lead malfunction. A total of 138 transvenous leads were analyzed, including 88 with known malfunction and 50 normal leads. The amplitude of a bipolar pacing stimulus on the ECG was recorded and compared with a control data set of newly implanted leads with bipolar stimuli normalized for output. The malfunction group consisted of 61% right atrium and 39% right ventricle leads with mean pacing output of 2.74 V at 0.5 ms. There was a significant difference in ECG bipolar stimulus amplitudes at time of identification of failure (7.89 ± 7.56 mm/V; P < .001) compared with those of normal leads (0.86 ± 0.41 mm/V). Receiver operating characteristic curve for the prediction of lead malfunction based on absolute ECG amplitude displayed an area under the curve of 0.93 (95% CI, 0.891–0.969). When normalized for programmed stimulus output, a cutoff of 5 mm/V demonstrated a sensitivity of 91% and a specificity of 92% (area under the curve, 0.967; 95% CI, 0.938–0.996). The maximum amplitude of a bipolar pacing stimulus on the ECG is significantly lower in normal functioning leads compared with those with known malfunction. This simply derived variable demonstrated good accuracy at identifying lead failure due to insulation breach.
ISSN:1547-5271
1556-3871
1556-3871
DOI:10.1016/j.hrthm.2024.03.1814