We report on the use of a commercially available band-edge detection system for substrate temperature monitoring of gallium arsenide substrates. The extension of the technique to the cases where strong absorption by either the substrate or substrate holder might normally preclude the use of such sys...
Gespeichert in:
Veröffentlicht in: | Journal of crystal growth 2007-04, Vol.301, p.88-92 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report on the use of a commercially available band-edge detection system for substrate temperature monitoring of gallium arsenide substrates. The extension of the technique to the cases where strong absorption by either the substrate or substrate holder might normally preclude the use of such systems due to poor signal levels is discussed. For indium-mounted wafers, a background subtraction/removal is applied which allows unambiguous determination of the band edge across the full temperature range. An alternative method of operation of the instrument as a highly configurable pyrometer allows measurements to be made on highly conducting p-type substrates where free carrier absorption swamps the band edge. |
---|---|