A new standard for sterility testing for autoclaved surgical trays

Sterility testing on autoclaved surgical trays provides scientific information on the duration for which an item can be safely stored. Previous studies were performed by sampling at various time-points after storage and did not take into account potential injuries (moving, lifting etc.) to these tra...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of hospital infection 1992-08, Vol.21 (4), p.253-260
Hauptverfasser: Widmer, A.F., Houston, A., Bollinger, E., Wenzel, R.P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Sterility testing on autoclaved surgical trays provides scientific information on the duration for which an item can be safely stored. Previous studies were performed by sampling at various time-points after storage and did not take into account potential injuries (moving, lifting etc.) to these trays as they occur in a clinical setting. A standardized protocol was developed by an expert panel based on events which were frequently observed in the Central Sterilization area. We tested 600 samples in 100 surgical trays that were moved and transported according to the standardized protocol. Moved surgical trays were contaminated in 8·3% of tests, a figure much higher than that reported by previous studies. These data suggest that current standards for sterility testing of autoclaved surgical trays can be improved by the addition of event-related testing.
ISSN:0195-6701
1532-2939
DOI:10.1016/0195-6701(92)90136-A