Traceability of Surface Temperature Measurements Using Contact Thermometers
Two methods of transferring traceability from contact thermometers calibrated by immersion to thermocouples measuring surface temperature are compared. The temperature of a reference surface is determined (i) by extrapolation from thermometers inserted at different depths, or (ii) by measurement on...
Gespeichert in:
1. Verfasser: | |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Two methods of transferring traceability from contact thermometers calibrated by immersion to thermocouples measuring surface temperature are compared. The temperature of a reference surface is determined (i) by extrapolation from thermometers inserted at different depths, or (ii) by measurement on the surface using a thermally compensated thermocouple. Components of measurement uncertainty are discussed for each technique, and results are compared from 50 deg C to 300 deg C. |
---|---|
ISSN: | 0094-243X |
DOI: | 10.1063/1.1627182 |