Crowd control of ions in the Astral analyzer

Space charge effects are the Achilles' heel of all high‐resolution ion optical devices. In time‐of‐flight mass analyzers, these may manifest as reduction of resolving power, mass measurement shift, peak coalescence, and/or transmission losses, while highly sensitive modern ion sources and injec...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of mass spectrometry. 2024-04, Vol.59 (4), p.e5006-n/a
Hauptverfasser: Stewart, Hamish, Grinfeld, Dmitry, Petzoldt, Johannes, Hagedorn, Bernd, Skoblin, Michael, Makarov, Alexander, Hock, Christian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Space charge effects are the Achilles' heel of all high‐resolution ion optical devices. In time‐of‐flight mass analyzers, these may manifest as reduction of resolving power, mass measurement shift, peak coalescence, and/or transmission losses, while highly sensitive modern ion sources and injection devices ensure that such limits are easily exceeded. Space charge effects have been investigated, by experiment and simulation study, for the astral multi‐reflection analyzer, incorporating ion focusing via a pair of converging ion mirrors, and fed by a pulsed extraction ion trap. Major factors were identified as the resonant effect between ~103 ions of similar m/z in‐flight and the expansion of trapped packets of ~104–5 ions prior to extraction. Optimum operation and compensated ion mirror calibration strategies were then generated and described based on these findings.
ISSN:1076-5174
1096-9888