Superior Dose and Energy Monitoring Capability of the Therma-Probe System

The repeatability of the Therma-Probe(R)-XP TP630XP series ion implant monitoring system is investigated and the implications on implant monitoring capabilities of the system discussed. Improvements in tool stability enable 10-day average dose detectability of 2% 3 using conventional on-the-fly meas...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Bakshi, M, Shaughnessy, D, Nicolaides, L, Mitchell, P
Format: Tagungsbericht
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The repeatability of the Therma-Probe(R)-XP TP630XP series ion implant monitoring system is investigated and the implications on implant monitoring capabilities of the system discussed. Improvements in tool stability enable 10-day average dose detectability of 2% 3 using conventional on-the-fly measurements, and 0.6% 3 using a recently introduced spatial averaging measurement mode. Dose detectability is the tool's figure of merit that combines both tool sensitivity and stability in a single parameter defining the noise-limited dose resolution of an ion implant metrology tool. The superior dose resolution of the Therma-Probe(R)-XP system enables implementation of in-line statistical process control for ion implant processes.
ISSN:0094-243X
DOI:10.1063/1.2401582