Impact of weight adjusted high frequency low tidal volume ventilation and atrial pacing in lesion metrics in high‐power short‐duration ablation: Results of a pilot study
Introduction Lesion size index (LSI) was introduced with the use of Tacticath™ and as a surrogate of lesion quality. The metric used to achieve the predetermined values involves combined information of contact force (CF), power and radiofrequency time. Rapid atrial pacing (RAP) and high‐frequency lo...
Gespeichert in:
Veröffentlicht in: | Journal of cardiovascular electrophysiology 2024-05, Vol.35 (5), p.975-983 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
Lesion size index (LSI) was introduced with the use of Tacticath™ and as a surrogate of lesion quality. The metric used to achieve the predetermined values involves combined information of contact force (CF), power and radiofrequency time. Rapid atrial pacing (RAP) and high‐frequency low‐tidal volume ventilation (HFLTV) independently or in combination improve catheter stability and CF and quality of lesions. Data of the impact of body weight adjusted HFLTV ventilation strategy associated with RAP in the lesion metrics still lacking. The study aimed to compare the results of high‐power short‐duration (HPSD) atrial fibrillation ablation using simultaneous weight adjusted HFLTV and RAP and standard ventilation (SV) protocol.
Methods
Prospective, nonrandomized study with 136 patients undergoing de novo ablation divided into two groups; 70 in RAP (100 ppm) + HFLTV with 4 mL/kg of tidal volume and 25 breaths/min (group A) and 66 patients with SV in intrinsic sinus rhythm (group B). Ablation using 50 W, CF of 5–10 g/10–20 g and 40 mL/minute flow rate on the posterior and anterior left atrial wall, respectively.
Results
No procedure‐related complications. Group A: Mean LSI points 70 ± 16.5, mean total lower LSI 3.4 ± 0.5, mean total higher LSI 8.2 ± 0.4 and mean total LSI 5.6 ± 0.6. Anterior and posterior wall mean total LSI was 6.0 ± 0.4 and 4.2 ± 0.3, respectively. Mean local impedance drop (LID) points were 118.8 ± 28.4, mean LID index (%) 12.9 ± 1.5, and mean LID |
---|---|