Validation of an online dual-loop cleanup device with an electrospray ionization tandem mass spectrometry-based system for simultaneous quantitative analysis of urinary benzene exposure biomarkers trans, trans-muconic acid and S-phenylmercapturic acid

The aim of this study is to validate isotope-dilution electrospray ionization tandem mass spectrometry (ESI–MS–MS) method with a dual-loop cleanup device for simultaneous quantitation of two benzene metabolites, trans, trans-muconic acid ( ttMA) and S-phenylmercapturic acid (SPMA), in human urine. I...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytica chimica acta 2006-01, Vol.555 (1), p.34-40
Hauptverfasser: Lin, Lung-Cheng, Chiung, Yin-Mei, Shih, Jeng-Fu, Shih, Tung-Sheng, Liao, Pao-Chi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of this study is to validate isotope-dilution electrospray ionization tandem mass spectrometry (ESI–MS–MS) method with a dual-loop cleanup device for simultaneous quantitation of two benzene metabolites, trans, trans-muconic acid ( ttMA) and S-phenylmercapturic acid (SPMA), in human urine. In this study, a pooled blank urine matrix from rural residents was adopted for validation of the analytical method. The calibration curve, detection limit, recovery, precision, accuracy and the stability of sample storage for the system have been characterized. Calibration plots of ttMA and SPMA standards spiked into two kinds of urine matrixes over a wide concentration range, 1/32–8-fold biological exposure indices (BEIs) values, showed good linearity ( R > 0.9992). The detection limits in pooled urine matrix for ttMA and SPMA were 1.27 and 0.042 μg g −1 creatinine, respectively. For both of ttMA and SPMA, the intra- and inter-day precision values were considered acceptable well below 25% at the various spiked concentrations. The intra- and inter-day apparent recovery values were also considered acceptable (apparent recovery >90%). The ttMA accuracy was estimated by urinary standard reference material (SRM). The accuracy reported in terms of relative error (RE) was 5.0 ± 2.0% ( n = 3). The stability of sample storage at 4 or −20 °C were assessed. Urinary ttMA and SPMA were found to be stable for at least 8 weeks when stored at 4 or −20 °C. In addition, urine samples from different benzene exposure groups were collected and measured in this system. Without tedious manual sample preparation procedure, the analytical system was able to quantify simultaneously ttMA and SPMA in less than 20 min.
ISSN:0003-2670
1873-4324
DOI:10.1016/j.aca.2005.08.075