Assessment of the Learning Curve for a Single-Use Disposable Duodenoscope

Background and Aims In response to documented duodenoscope-related infectious outbreaks of multidrug-resistant organisms, the Food and Drug Administration has recommended a transition to duodenoscopes with innovative designs, including duodenoscopes with disposable components or fully disposable duo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Digestive diseases and sciences 2024-06, Vol.69 (6), p.1956-1962
Hauptverfasser: Saleem, Nasir, Tong, Yan, Sherman, Stuart, Gromski, Mark A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background and Aims In response to documented duodenoscope-related infectious outbreaks of multidrug-resistant organisms, the Food and Drug Administration has recommended a transition to duodenoscopes with innovative designs, including duodenoscopes with disposable components or fully disposable duodenoscopes. We aim to characterize the learning curve (LC) for a single-use disposable duodenoscope. Methods We performed a retrospective analysis of a prospectively collected database from 31 patients who underwent ERCP by a single, experienced operator using the EXALT Model D® (Boston Scientific, Marlborough) disposable duodenoscope at a single tertiary referral center. The LC for this device was described by the number of cases needed to achieve proficiency using cumulative sum (CUSUM) analysis. Number of attempts to cannulate and time to cannulate the desired duct were assessed as separate endpoints. The overall mean number of attempts and overall mean time to cannulation were used as the target values in the respective CUSUM analyses. Proficiency was defined as the number of procedures where an inflection point was reached in the CUSUM graph. This observation indicates improving operator performance as shown by a decrease in the number of attempts and shortening of cannulation time after the defined number of procedures. Results Overall, 31 patients underwent ERCP using the EXALT Model D disposable duodenoscope by a single experienced endoscopist. 6 (19%) patients had a native papilla and the majority of these procedures were classified as ASGE complexity level 2 or above. The procedure was completed using solely the disposable duodenoscope in 27 patients (87%), while a reusable duodenoscope was required for procedure completion in 4 patients (13%). The cross-overs were distributed evenly across the performance period. Procedure-related adverse events included: post-ERCP pancreatitis (3%), bleeding (3%) and no perforations. In the analyses of both endpoints, an inflection of the CUSUM curves is achieved at 10 cases, indicating sustained reduction of cannulation attempts and time to cannulation. Conclusion Among experienced pancreaticobiliary endoscopists, approximately 10 ERCPs is the threshold whereby procedure-related factors including cannulation success and procedural time improves. Procedure-related adverse events are consistent with those expected with reusable duodenoscopes. The need to cross-over from single-use duodenoscope to reusable duodenoscope
ISSN:0163-2116
1573-2568
1573-2568
DOI:10.1007/s10620-024-08305-z