A Cost-Effective Approach to Resistant AV Fistula Stenosis: Successful Treatment Using Coronary OPN NC® Balloon in a Low-Income Setting

Stenosis in the anastomotic site or venous limb of an arteriovenous fistula (AVF) is the most frequent cause of AVF failure. Percutaneous angioplasty with a standard or high-pressure balloon is the first-line treatment for AVF stenosis due to its higher technical success rate (90%) and lower complic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Seminars in dialysis 2024-05, Vol.37 (3), p.269-272
Hauptverfasser: Lalani, Kanhai, Rao, M Sudhakar, Sagar, M Harsha, R, Padmakumar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Stenosis in the anastomotic site or venous limb of an arteriovenous fistula (AVF) is the most frequent cause of AVF failure. Percutaneous angioplasty with a standard or high-pressure balloon is the first-line treatment for AVF stenosis due to its higher technical success rate (90%) and lower complication rate (4%). Almost 20% of stenosis cases are resistant or undilatable by regular-pressure balloon angioplasty due to fibrosis, leading to technical failure or restenosis. Alternative therapies, such as atherectomy devices or cutting balloons, are expensive and difficult to obtain in low-income developing countries. We successfully treated resistant AVF stenosis with a coronary OPN-NC® ultra-high-pressure balloon and produced a good angiographic result with technical success. Coronary hardware is easily available and relatively cheaper compared to dedicated peripheral balloons or devices in our country due to reuse, which can be a boon in such type of cases. According to the standard hospital protocol, Cathlab hardware was reused.
ISSN:0894-0959
1525-139X
DOI:10.1111/sdi.13196