Mapping of uranium particles on J-type swipes with microextraction-ICP-MS

A microextraction liquid sampling system coupled to a quadrupole inductively coupled plasma-mass spectrometer (ICP-MS) was utilized to spatially discern uranium particles, isotopically, on a cellulose-based swipe material ( i.e. , J-type swipe). These types of swipes are often used by the Internatio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analyst (London) 2024-04, Vol.149 (8), p.2244-2251
Hauptverfasser: Bradley, Veronica C, Burleson, Jacob, Andrews, Hunter B, Thompson, Cyril V, Spano, Tyler L, Dunlap, Daniel R, Zirakparvar, N. Alex, Ticknor, Brian W, Hexel, Cole R, Manard, Benjamin T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A microextraction liquid sampling system coupled to a quadrupole inductively coupled plasma-mass spectrometer (ICP-MS) was utilized to spatially discern uranium particles, isotopically, on a cellulose-based swipe material ( i.e. , J-type swipe). These types of swipes are often used by the International Atomic Energy Agency (IAEA) as part of their environmental sampling program. A grid was created such that extraction locations covered the center circle ( n = 34 without overlapping). Uranium (U) particulates (