High-performance liquid chromatographic assay of phosphate and organophosphorus pesticides using a post-column photochemical reaction and fluorimetric detection
A sensitive method for the post-column reaction detection of organophosphorus pesticides is described. The method relies on photolysis of the organocompounds by irradiation with a low-pressure mercury lamp (main spectral line, 254 nm) in the presence of peroxydisulfate. The resultant orthophosphate...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2005-06, Vol.540 (2), p.383-391 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A sensitive method for the post-column reaction detection of organophosphorus pesticides is described. The method relies on photolysis of the organocompounds by irradiation with a low-pressure mercury lamp (main spectral line, 254
nm) in the presence of peroxydisulfate. The resultant orthophosphate was reacted with molybdate to form molybdophosphoric acid, which subsequently reacted with thiamine to generate thiochrome. Finally, the fluorescence intensity of thiochrome was measured at 440
nm with excitation at 375
nm. Factors affecting the rate of these reactions were optimized so that its contribution to the total band-broadening was negligible.
This detection system was used for the determination of phosphate, acephate and methamidophos, which were separated on an ODS column by isocratic reversed phase chromatography with acetonitrile–water as the mobile phase. A linear relationship between analyte concentration and peak area was obtained within the range 0.016–7.0
μg
ml
−1 with correlation coefficients greater than 0.9995 and detection limits between 4 and 12
ng
ml
−1. Intra- and inter-day precision values of about 1.2% R.S.D. (
n
=
10) and 2.1% R.S.D. (
n
=
30), respectively, were obtained.
Pesticide residues below ng
ml
−1 levels could be determined in environmental waters when a preconcentration device was coupled on-line with the HPLC system. Detection limits as low as 0.01
ng
ml
−1 were achieved for only 250
ml of sample. In the analyses of vegetables and grains, the detection limit was about 1
μg
kg
−1. |
---|---|